Table 3a: Within-batch precision and accuracy of the method for determining FAV in plasma samples.
|
Batch No: 1 (n = 6) |
Batch No: 2 (n = 6) |
Batch No: 3 (n = 6) |
||||||
Nominal Conc. (ng/mL) |
Mean Conc. Found (ng/mL) |
RE (%) |
CV (%) |
Mean Conc. Found (ng/mL) |
RE (%) |
CV (%) |
Mean Conc. Found (ng/mL) |
RE (%) |
CV (%) |
80 (LLOQ) |
86.418 |
8.022 |
2.582 |
80.918 |
1.147 |
2.103 |
78.420 |
-1.975 |
4.404 |
240(QC Low) |
224.505 |
-6.456 |
1.580 |
222.676 |
-7.218 |
1.870 |
217.154 |
-9.519 |
2.476 |
1500(QC Medium1) |
1389.945 |
-7.337 |
1.279 |
1399.869 |
-6.675 |
2.372 |
1352.137 |
-9.858 |
2.151 |
12000 (QC Medium2) |
11760.785 |
-1.993 |
1.927 |
11678.196 |
-2.682 |
2.094 |
11650.015 |
-2.917 |
1.999 |
24000(QC High) |
23745.423 |
-1.061 |
0.964 |
24099.378 |
0.414 |
1.214 |
23361.479 |
-2.661 |
1.928 |