Table 4: Comparison between experimental and simulated coincidence summing correction factors for Marinelli beaker source.
Volume (ml) | Energy (keV) | CSFExp | CSFCal | CSFSimu | RD (%) |
450 | 1173.24 | 0.875 | 0.901 | 0.880 | -0.5 |
1332.50 | 0.858 | 0.900 | 0.867 | -0.1 | |
898.02 | 0.909 | ||||
1836.01 | 0.847 | ||||
600 | 1173.24 | 0.874 | 0.906 | 0.882 | -0.9 |
1332.50 | 0.867 | 0.905 | 0.871 | -0.4 | |
898.02 | 0.901 | ||||
1836.01 | 0.831 | ||||
800 | 1173.24 | 0.872 | 0.909 | 0.886 | -1.6 |
1332.50 | 0.868 | 0.908 | 0.875 | -0.8 | |
898.02 | 0.896 | ||||
1836.01 | 0.824 | ||||
1000 | 1173.24 | 0.884 | 0.911 | 0.889 | -0.5 |
1332.50 | 0.877 | 0.910 | 0.891 | -1.5 | |
898.02 | 0.893 | ||||
1836.01 | 0.812 |