Table 2: The measurement errors due to misalignment of components.

Errors due to misalignment of the polarizer

Orientation (°)

85

86

87

88

89

90

91

92

93

94

95

δ(°)

38.3

39.6

40.9

42.2

43.6

45

46.4

47.9

49.4

51.0

52.5

θ(°)

15.0

15.0

15.0

15.0

15.0

15.0

15.0

15.0

15.0

15.0

15.0

Errors due to misalignment of the 1st waveplate

Orientation (°)

-50

-49

-48

-47

-46

-45

-44

-43

-42

-41

-40

δ(°)

37.8

38.9

40.3

41.7

43.3

45.0

46.8

48.6

50.5

52.5

54.5

θ(°)

24.0

21.9

20.0

18.2

16.6

15.0

13.6

12.2

10.9

9.8

8.6

Errors due to misalignment of the 2nd waveplate

Orientation (°)

-50

-49

-48

-47

-46

-45

-44

-43

-42

-41

-40

δ(°)

37.8

38.9

40.3

41.7

43.3

45.0

46.8

48.6

50.5

52.5

54.5

θ(°)

20.2

18.9

17.7

16.7

15.8

15.0

14.4

13.8

13.4

13.1

12.9

Errors due to misalignment of the analyzer

Orientation (°)

85

86

87

88

89

90

91

92

93

94

95

δ(°)

53.8

52.0

50.3

48.5

46.7

45.0

43.3

41.6

39. 9

38.2

36.6

θ(°)

11.9

12.5

13.1

13.7

14.3

15.0

15.7

16.5

17.3

18.2

19.2